SENSe Subsystem R&S FSL
1300.2519.12 6.190 E-11
[SENSe<1|2>:]ESPectrum:PRESet:RESTore
This command copies the XML files from the C:\R_S\instr\sem_backup folder to the
C:\R_S\instr\sem_std folder. Files of the same name are overwritten.
The numeric suffixes <1|2> are not relevant.
This command is an event and therefore has no *RST value and no query.
This command is available from firmware version 1.80.
Example
ESP:PRES:REST
Restores the originally provided XML files.
Characteristics
RST value:
SCPI: device–specific
Mode
A, CDMA, EVDO
[SENSe<1|2>:]ESPectrum:PRESet:STORe
This command saves the specified XML file under C:\r_s\instr\sem_std. To save the file in
subdirectory, include the relative path.
The numeric suffixes <1|2> are not relevant.
This command is an event and therefore has no *RST value and no query.
Example
ESP:STOR 'DL\BandClass_20.xml'
Saves the BandClass_20.xml XML file in the C:\R_S\instr\sem_std\DL directory.
Characteristics
RST value:
SCPI: device–specific
Mode
A, CDMA, EVDO
[SENSe<1|2>:]ESPectrum:RANGe<1...20>:BANDwidth /
[SENSe<1|2>:]ESPectrum:RANGe<1...20>:BANDwidth:RESolution
This command sets the RBW value for the specified range.
The numeric suffixes <1...20> specify the range. The numeric suffixes <1|2> are not relevant.
This command is available from firmware version 1.80.
Parameter
Refer to the data sheet.
Example
ESP:RANG2:BAND:RES 5000
Sets the RBW for range 2 to 5 kHz.