SENSe Subsystem R&S FSL
1300.2519.12 6.194 E-11
Example
ESP:RANG3:INP:ATT 10
Sets the attenuation of range 3 to 10 dB.
Characteristics
RST value: 0 dB
SCPI: device–specific
Mode
A, CDMA, EVDO
[SENSe<1|2>:]ESPectrum:RANGe<1...20>:INPut:ATTenuation:AUTO
This command activates or deactivates the automatic RF attenuation setting for the specified
range.
The numeric suffixes <1...20> specify the range. The numeric suffixes <1|2> are not relevant.
This command is available from firmware version 1.80.
Parameter
ON | OFF
Example
ESP:RANG2:INP:ATT:AUTO OFF
Deactivates the RF attenuation auto mode for range 2.
Characteristics
RST value: ON
SCPI: device–specific
Mode
A, CDMA, EVDO
[SENSe<1|2>:]ESPectrum:RANGe<1...20>:INPut:GAIN:STATe
This command switches the preamplifier on or off for the specified range.
The numeric suffixes <1...20> specify the range. The numeric suffixes <1|2> are not relevant.
This command is available from firmware version 1.80.
Parameter
ON | OFF
Example
ESP:RANG3:INP:GAIN:STATe ON
Switches the preamplifier for range 3 on or off.
Characteristics
RST value: OFF
SCPI: device–specific
Mode
A, CDMA, EVDO