SENSe Subsystem R&S FSL
1300.2519.12 6.196 E-11
Example
ESP:RANG1:LIM:ABS:STOP 20
Sets an absolute limit of 20 dBm at the stop frequency of the range.
Characteristics
RST value: –13 dBm
SCPI: device–specific
Mode
A, CDMA, EVDO
[SENSe<1|2>:]ESPectrum:RANGe<1...20>:LIMit:RELative:STARt
This command sets a relative limit value at the start frequency of the specified range. Different
from manual operation, this setting is independently of the defined limit check type.
The numeric suffixes <1...20> specify the range. The numeric suffixes <1|2> are not relevant.
This command is available from firmware version 1.80.
Parameter
–400 to in 400 dBc
Example
ESP:RANG3:LIM:REL:STAR –20
Sets a relative limit of –20 dBc at the start frequency of the range.
Characteristics
RST value: –50 dBc
SCPI: device–specific
Mode
A, CDMA, EVDO
[SENSe<1|2>:]ESPectrum:RANGe<1...20>:LIMit:RELative:STOP
This command sets a relative limit value at the stop frequency of the specified range. Different
from manual operation, this setting is independently of the defined limit check type.
The numeric suffixes <1...20> specify the range. The numeric suffixes <1|2> are not relevant.
This command is available from firmware version 1.80.
Parameter
–400 to in 400 dBc
Example
ESP:RANG3:LIM:REL:STOP 20
Sets a relative limit of 20 dBc at the stop frequency of the range.
Characteristics
RST value: –50 dBc
SCPI: device–specific
Mode
A, CDMA, EVDO