SENSe Subsystem R&S FSL
1300.2519.12 6.198 E-11
[
SENSe<1|2>:]ESPectrum:RANGe<1...20>:SWEep:TIME
This command sets the sweep time for the specified range.
T
he numeric suffixes <1...20> specify the range. The numeric suffixes <1|2> are not relevant.
This command is available from firmware version 1.80.
Parameter
Allowed values depend on the ratio of span to RBW and RBW to VBW. For details refer to the
data sheet.
Example
ESP:RANG1:SWE:TIME 1
Sets the sweep time for range 1 to 1 s.
Characteristics
RST value: 0.27 s
SCPI: device–specific
Mode
A, CDMA, EVDO
[SENSe<1|2>:]ESPectrum:RANGe<1...20>:SWEep:TIME:AUTO
This command activates or deactivates the automatic sweep time setting for the specified
range.
The numeric suffixes <1...20> specify the range. The numeric suffixes <1|2> are not relevant.
This command is available from firmware version 1.80.
Parameter
ON | OFF
Example
ESP:RANG3:SWE:TIME:AUTO OFF
Deactivates the sweep time auto mode for range 3.
Characteristics
RST value: ON
SCPI: device–specific
Mode
A, CDMA, EVDO
[SENSe<1|2>:]ESPectrum:RANGe<1...20>:TRANsducer
This command sets a transducer for the specified range. You can only choose a transducer that
fulfills the following conditions:
The transducer overlaps or equals the span of the range.
The x–axis is linear.
The unit is dB.
The numeric suffixes <1...20> specify the range. The numeric suffixes <1|2> are not relevant.
This command is available from firmware version 1.80.
Parameter