SENSe Subsystem R&S FSL
1300.2519.12 6.218 E-11
[
SENSe<1|2>:]MPOWer:FTYPe
This command defines the filter type for the measurement.
T
he numeric suffixes <1|2> are not relevant.
Parameter
NORMal | CFILter | RRC
Example
SENSe:MPOWer:FTYPe CFILter
Characteristics
*RST value: –
SCPI: device–specific
Mode
A
[SENSe<1|2>:]MPOWer:RESult[:LIST]?
This command queries the results of a multiple burst power measurement as configured and
initiated with [SENSe<1|2>:]MPOWer[:SEQuence]. The results are output in a comma–
separated list of floating point values. The unit used for the return values is always dBm.
This command may be used to obtain measurement results in an asynchronous way using the
service request mechanism for synchronization with the end of the measurement.
The numeric suffixes <1|2> are not relevant.
If no measurement results are available, the command will return a query error.
Example
*ESE 1
*SRE 32
Configuration of status reporting systems for the generation of an SRQ on operation complete
SENSe:MPOWer 935.2MHz,1MHz,434us,VIDEO,50PCT,5us,MEAN,20;*OPC
Configuring and starting the measurement
...
Further actions of the control computer during measurement
On SRQ:
Response to service request
SENSe:MPOWer:RESult?
Characteristics
*RST value: –
SCPI: device–specific
Mode
A