R&S FSL SENSe Subsystem
1300.2519.12 6.219 E-11
[
SENSe<1|2>:]MPOWer:RESult:MIN?
This command queries the minimum power value in a multiple burst power measurement as
configured and initiated with [SENSe<1|2>:]MPOWer[:SEQuence]. The unit used for the
return values is always dBm.
T
he numeric suffixes <1|2> are not relevant.
If no measurement result is available, the command will return a query error.
Example
*ESE 1
*SRE 32
Configuration of status reporting systems for the generation of an SRQ on operation complete
SENSe:MPOWer 935.2MHz,1MHz,434us,VIDEO,50PCT,5us,MEAN,20;*OPC
Configuring and starting the measurement
...
Further actions of the control computer during measurement
On SRQ:
Response to service request
SENSe:MPOWer:RESult:MIN?
Characteristics
*RST value: –
SCPI: device–specific
Mode
A