EasyManuals Logo

YOKOGAWA DL750 User Manual

YOKOGAWA DL750
747 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #119 background imageLoading...
Page #119 background image
3-16
IM 701210-05E
• The BNC part of the Passive Probe for the DL750/DL750P (701940) is made
of metal. Therefore, use the probe at 42 V or less for isolated inputs (701250
(HS10M12), 701251 (HS1M16), 701260 (HV (with RMS)), 701275 (ACCL/
VOLT), 701280 (FREQ), etc.) for safety reasons. (Do not connect voltage
above 42 V to both the High and Low sides.) For non-isolated inputs (701255
(NONISO_10M12), etc.), fasten the module screws.
• The measurement category of the 701260 (HV (with RMS)) is 400V-CATII on
the low side and 700V-CATII on the high side. Use caution because the
overvoltage tolerance differs between the low and high sides.
• When applying high voltage to the 701280 (FREQ), use the isolated probe
(700929).
• In Using the High Voltage Differential Probes
Be sure to connect the GND lead of the differential probe (700924/700925) to
the functional ground terminal on the right side panel of the DL750/DL750P.
Otherwise, high voltage may appear at the BNC connector making it dangerous.
• Applying a voltage exceeding the value indicated below may damage the input
section. If the frequency is above 1 kHz, damage may occur even when the
voltage is below this value.
• High-Speed 10 MS/s, 12-Bit Isolation Module (701250)
Maximum input voltage (at a frequency of 1 kHz or less)
• Combined with the isolated probe 700929 (10:1)
1
600 V (DC+ACpeak)
• Safety cable (1:1) (combined with 701901+701954)
5
or direct input
9
250 V (DC+ACpeak)
Maximum allowable common mode voltage (at a frequency of 1 kHz or
less)
• Combined with the isolated probe 700929 (10:1)
2
or safety cable (1:1)
(combined with 701901+701954)
8
400 Vrms (CAT I), 300 Vrms (CAT II)
• Direct input
10
42 V (DC+ACpeak) (CAT I and CAT II, 30 Vrms)
• High-Speed High-Resolution 1 MS/s, 16-Bit Isolation Module (701251)
Maximum input voltage (at a frequency of 1 kHz or less)
• Combined with the isolated probe 700929 (10:1)
1
600 V (DC+ACpeak)
• Safety cable (1:1) (combined with 701901+701954)
5
or direct input
9
140 V (DC+ACpeak)
Maximum allowable common mode voltage (at a frequency of 1 kHz or less)
• Combined with the isolated probe 700929 (10:1)
2
or safety cable (1:1)
(combined with 701901+701954)
8
400 Vrms (CAT I), 300 Vrms (CAT II)
• Direct input
10
42 V (DC+ACpeak) (CAT I and CAT II, 30 Vrms)
• High-Speed 10 MS/s, 12-Bit Non-Isolation Module (701255)
This module is non-isolated. Be sure to fasten the module screws when
measuring a voltage above 42 V on this module. In addition, use the
dedicated non-isolated passive probe (10:1) (701940).
Maximum input voltage (at a frequency of 1 kHz or less)
• Combined with the passive probe (701940) (10:1)
600 V (DC+ACpeak)
• Direct input
9
250 V (DC+ACpeak)
3.6 Connecting Probes

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL750 and is the answer not in the manual?

YOKOGAWA DL750 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL750
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals