EasyManua.ls Logo

YOKOGAWA DL750 - Page 47

YOKOGAWA DL750
747 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2-9
IM 701210-05E
Explanation of Functions
2
2.2 Setting the Horizontal and Vertical Axes
Probe Type <Section 5.6>
For voltage (current) measurement, a probe is normally used in connecting the circuit
being measured to the measurement input terminal. Using a probe has the following
advantages.
Prevents disturbing the voltage and current of the circuit being measured.
Inputs the signal with no distortion.
Expands the voltage range that can be measured by the DL750/DL750P.
When using a probe, the attenuation or current-to-voltage conversion ratio setting on the
module must be set equal to the probe attenuation on the DL750/DL750P so that the
measured voltage can be read directly.
The probe attenuation of each probe (accessory sold separately) is set as follows:
Voltage probe (700929): 10:1
Current probe (700937/701933): 10 A:1 V
Current probe (701930/701931): 100 A:1 V
Voltage differential probe (700924): 1000:1, 100:1
10:1 passive probe for the DL750/DL750P: 10:1
The DL750/DL750P has the following attenuation settings: 1:1, 10:1, 100:1, 1000:1, 10
A:1 V
1
, and 100 A :1 V
2
. If you are using a probe other than the ones provided as
accessories (sold separately), set the attenuation ratio on the DL750/DL750P according
to the attenuation of the probe.
1. Output voltage rate: 0.1 V/A
2. Output voltage rate: 0.01 V/A
Note
Use a probe that matches the input capacity of each module. Otherwise, the capacity cannot
be adjusted.
Bandwidth Limit <Section 5.7>
The bandwidth can be limited on each module. By limiting the bandwidth, you can
eliminate the noise components from the input signal for waveform observation.

Table of Contents

Related product manuals