EasyManuals Logo

YOKOGAWA DL750 User Manual

YOKOGAWA DL750
747 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #570 background imageLoading...
Page #570 background image
16-18
IM 701210-06E
Periodic Mail
The DL750/DL750P status can be sent periodically to a specified mail address on the
network.
• DL750/DL750P Information That Is Transmitted
• Acquisition Condition
Start/Stop, trigger condition, and acquisition counter
• Error Log Information
Error number and corrective action in English (up to 16 errors)
• Success/Fail Information of GO/NO-GO
Only when GO/NO-GO determination is executed (see sections 11.8 and 11.9)
• Results of Waveform Parameter Measurement
Only when automated measurement of waveform parameters (see section 11.6) or
GO/NO-GO determination using waveform parameters (see section 11.8) is
executed.
The maximum number of measurement results of waveform parameters that are
sent is the same as the maximum number (24) of measurement results displayed
on the DL750/DL750P.
• Sample Periodic Mail
[Comment] aaaaaa
[ACQ Status] Stopped 162
[GO/NOGO Status] Success: 140 Fail: 21
Max (C1) 4.16667V
SDv (C2) 697.941mV
Freq(C3) 500.0000kHz
+Wd (C4) 1.00us>
ErrNo 004 Exit from GO/NO-GO mode.
ErrNo 806 Cannot change settings during GO/NO-GO.
Stop the GO/NO-GO.
ErrNo 004 Exit from GO/NO-GO mode.
ErrNo 806 Cannot change settings during GO/NO-GO.
Stop the GO/NO-GO.
ErrNo 602 No SCSI device or no media inserted.
Check the SCSI device connection and
the SCSI ID, and make sure that the storage
medium is inserted (if applicable).
16.5 Sending E-mail Messages (Periodic Mail or Action Mail)