IM 701210-06E
Measurement Range
You can set the range on the time axis for measuring the waveform parameter.
Range1 and Range2 are the start and end points, respectively.
Selectable range: ±5 div
Resolution: Varies depending on the record length.
However, the measurement start point cannot be set to the same point as the
measurement end point or to a value to the right of the measurement end point.
Turning ON/OFF the Statistics Display
The statistics (Max, Min, and Average) of the waveform parameters retrieved
into the PC can be displayed.
To display the statistics, select the Statistics check box. Clear the check box to
not display the statistics.
The number of measurement values displayed in the cell of Excel is up to the
number specified in “Number of Displayed Measured Values (Data Number)” on
the previous page. The statistics are determined on all the measured data since
the trend display was started, not on only the displayed measured values.
Updating the Setup Condition
Click Update to update the Marker Position and Range that are set on the control
Panel of the measurement trend window to match the newest setup conditions of
the DL750/DL750P.
Starting/Stopping the Trend Display of the Measured Values of the Waveform
Parameters
Start
Click Start/Stop. The indicator illuminates in yellow, and retrieval of the measured
values of waveform parameters starts at the specified retrieve interval. At the same
time, Microsoft Excel starts, and the measured values and trend graphs are displayed/
drawn. If the statistical display is turned ON, the statistics are also displayed. The
retrieval of the measured value of waveform parameter continues until the operation is
stopped.
Stop
Click Start/Stop while the trend display is in progress. The indicator turns OFF,
and the display and drawing of the measured values and trend graph stop.
If you attempt to save the data to a file or close Excel while the trend display is in
progress, a runtime error occurs. In this case, select No on the runtime error
dialog box and close the dialog box. Then, click Start/Stop on the measurement
trend window to stop the trend display. After stopping the trend display, save the
file or close Excel.
16.7 Using the Web Server Function

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL750 and is the answer not in the manual?

YOKOGAWA DL750 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL750
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals