3. Repair Instructions 1/2010 3 - 95
BA-TE-DE08C M.KAY
Dialog+ SW9xx_sm_Chapter 3_1-2010.doc/pdf <100329> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
3.7 Default Table SW 9.xx
System Configuration
Unit HD HDF Online US Default Modified
Dialysate Side Parameters
Na Concentration/Conductivity Measurement Unit - mS/cm mS/cm mS/cm
Acetate Mode Disabled - NO YES YES
Default Conductivity Mode - BIC BIC BIC
Stand-by Function Available in Preparation - YES YES NO
Stand-by after Self Test/Rinsing - YES YES NO
Maximum Stand-by Time h:min 1:00 1:00 5:00
Selected Acetate -
B.BRAUN
CONCENTRATE 1
B.BRAUN
CONCENTRATE 1
B.BRAUN
CONCENTRATE 1
Selected Bicarbonate -
BIC without NaCl BIC without NaCl BIC without NaCl
Selected Acid -
B.BRAUN ACID 1 B.BRAUN ACID 1 B.BRAUN ACID 1
Temperature Measurement Unit - °C °C °C
Dialysate Flow
ml/min
500 500 600
Dialysate Temperature °C 37.0 37.0 37.0
Settings for Acetate Concentrate
Unit
1
st
to 10
th
Concentrate Name - B.Braun Concentrate 1 – 10
Conversion Factor mS/cm per mmol/l 0.10500
Default Value mS/cm 14.0
Settings for BIC
1
st
Bicarbonate without NaCl - BIC without NaCl
Conversion Factor mS/cm per mmol/l 0.0950
Default Value mS/cm 3.0
Acid Selection - B.Braun Acid 1
2
nd
Bicarbonate with NaCl - BIC with NaCl
Conversion Factor mS/cm per mmol/l 0.0950
Default Value mS/cm 5.8
Acid Selection - B.Braun Acid 2
Setting for Acid -
1
st
to 20
th
Concentrate Name - B.Braun Acid 1 – 20
Conversion Factor mS/cm per mmol/l 0.10500
Default Value mS/cm 14.3
Blood Side Parameters -
BS Pressure Test with Equalisation of Pressure - NO NO YES
Default Blood Flow at End of Therapy ml/min 100 100 200
AV Line without Chamber is Possible - NO NO NO
AV Line with PBE Connector - YES YES NO
Min-Max Parameters
Delta PA min. mmHg 70 70 70
Delta PA max. mmHg 70 70 70
Limit Maximum PBE mmHg 700 700 700
Limit Delta PBE mmHg 150 150 600
Limit Maximum TMP mmHg 350 350 500
Limits Low/High TMP - active active active
Low Limit TMP % 40 40 40
High Limit TMP % 40 40 40
Minimum TMP Alarm Window mmHg 20 20 20
Extended TMP Limit Range Button is Displayed in
Treatment
- NO NO NO
Select Extended TMP Limit Range - inactive inactive inactive
Arterial Bolus Parameters
Arterial Bolus Volume ml 100 100 100
Arterial Bolus with SAKA Support NO NO NO