4. TSM Service Program 1/2010 4 - 42
BA-TE-DE08C M.KAY Dialog+ SW9xx_SM_Chapter 4-1-1_1-2010.doc/pdf <110301> yymmdd B. Braun Avitum AG
Dialog
+
SW 9.xx
Disinfection Configuration Data
Chemical (long), Chemical (short)
The parameters for the disinfectants are set in the
Disinfection Configuration Data, Chemical (Long),
Chemical (Short) table. The following parameters
are available:
Disinfectant Name, Type and Volume
Reaction Time Long/Short
Rinsing Time
Operating Temperature
Minimum Conductivity
Details can be found in Chapter 3 repair
instructions.
Disinfection Configuration Data
Central Thermal, Central Chemical, Rinsing
The parameters for the water inlet system can be
set in the menu Disinfection Configuration Data,
Central Thermal, Central Chemical, Rinsing.
Central Thermal:
Inlet Flow
Time
Central Manual Chemical:
Inlet Flow
Inlet Time
Rinsing Flow
Rinsing Time
Central Automatic Chemical:
Inlet Volume
Switch-off for Retention without Automatic
Switch-on
Retention Time
Rinsing Flow
Rinsing Time
Rinsing:
Inlet Flow
Time