4. TSM Service Program 1/2010 4 - 116
BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd B. Braun Avitum AG
Dialog
+
SW 9.xx
A common calibration or a single calibration can
be performed for the pressure sensors.
Common Calibration
Press the
key for a combined
calibration of the pressure sensors PA, PV, PBE
and PBS. A calibration window is opened if the
key is selected. The common calibration of all
blood pressure sensors is recommended. The menu
shows the Common Calibration.
Single Calibration
Press an individual
key for a
single calibration of the pressure sensor PA, PV,
PBE or PBS. A calibration window is opened if the
key is selected.
Press the
key to enter a
reference value. A keypad is opened if the key is
selected.
Pay attention to the negative or
positive sign during the enter of the
reference value.
• Pressure measurement instrument (e.g. HDM 99 or equivalent)
• Syringe
• Stopcock system with lines (770203A) or tubing/tubing clamp
3. Connect the pressure measurement instrument, syringe and stopcock
system.
4. Connect the tubing to the respective pressure sensors PA, PV, PBE and/or
PBS (option).
5. Press
key.
PA/PV/PBE/PBS Calibration Point
-100 mmHg
6. Build up a pressure of -100 mmHg (+10 mmHg) with a syringe and clamp
the tube.
7. Press
key.
Wait until digits are stable.
8. Enter the stable value from the pressure measurement instrument as
reference value.
9. Confirm with the
key.