Brooks Automation 5. Operational Interfaces
Part Number: 605914 Rev. B Power Connections
Power Connections
TheMarathonExpressrequires200-240VACat15Amps(max)50-60Hzsinglephase(phase-to-
phaseorphase-to-neutral)forstandardoperation.Theelectricchamberheatingoptionrequires
200-240VACat15Amps(max)50-60Hz(phase-to-phaseorphase-to-neutral).Theactualpower
beingdrawnwilldependuponoperationsbeingperformed,howeverallpowerwiringmustbe
capableofcarryingthefullload,referto"FacilitiesSpecifications"onpage21.
SeetheControlsManualandSystemBlockDiagramformoreinformation.
NOTE: Thefacilityisresponsibleforthemaindisconnectdevicebetweenthe
MarathonExpressandthefacility’spowersource,ensuringitcomplieswiththe
correctelectriccodes.ServicetotheMXsystemshouldhavetheappropriatefuseor
circuitbreakerrating.Themainuser-installeddisconnectdevicemustberatedat
10,000AICminimum.Allcurrentrequirementsshownaremaximumvalues,theactual
currentdrawnwilldependupontheuseoftheMarathonExpress.
Theproperdistributionofpowerisdeterminedbytheuser.BrooksAutomationrecommendscircuit
breakers,ground-faultprotection,contactors,and/orEMOcircuitryforequipmentandpersonal
safety.
ThefacilityisresponsibleforthemaindisconnectdevicebetweentheMarathonExpressandthe
facility’spowersource,ensuringitcomplieswiththecorrectelectriccodes.ServicetotheMX
systemshouldhavetheappropriatefuseorcircuitbreakerrating.Allcurrentrequirementsshown
aremaximumvalues,theactualcurrentdrawnwilldependupontheuseoftheMarathonExpress
400.Refertothe"FacilitiesSpecifications"onpage21forpowerrequirements.
NOTE: BeforeconnectingordisconnectingmainpowertotheMarathonExpress,
ensureallcircuitbreakersontheMXsystemandanyrelatedequipmentareinthe
OFFposition.Damagetointernalcomponentsmayresultifcircuitbreakersareleftin
theONposition.
Copyright © 2023, Brooks Automation, Inc.
105