Product Specifications
II-8
Comen SpO
2
: 0% - 100%
Masimo SpO
2
: 1% - 100%
Nellcor SpO
2
: 0% - 100%
SpO
2
Measurement
accuracy
Comen SpO
2
:
within range 70% - 100%: ±2% in Adu./Ped. mode (non-motion),
±3% in Neo. mode (non-motion);
within range 0% - 69%: not defined.
Masimo SpO
2
:
with range 70% - 100%: ±2% in Adu./Ped. Mode (non-motion),
±3% in Adu./Ped. mode (during motion),
±3% in Neo. mode (motion and non-motion);
within range 1% - 69%: not defined.
Nellcor SpO
2
:
within range 70% - 100%: ±2% in Adu./Ped. mode (non-motion),
±3% in Neo. mode (non-motion);
within range 0% - 69%: not defined.
Comen SpO
2
Upper alarm limit: (lower limit+2%) - 100%;
Lower alarm limit: (extreme low limit+2%) - (upper limit-2%);
Extreme low alarm limit: 0% - (lower limit-2%)
Masimo SpO
2
Upper alarm limit: (lower limit+2%) - 100%;
Lower alarm limit: (extreme low limit+2%) - (upper limit-2%);
Extreme low alarm limit: 1% - (lower limit-2%)
Nellcor SpO
2
Upper alarm limit: (lower limit+2%) - 100%;
Lower alarm limit: (extreme low limit+2%) - (upper limit-2%);
Extreme low alarm limit: 0% - (lower limit-2%)
ΔSpO
2
Alarm limit setting
Alarm limit range: 0-50 %. Step: 1%.
PI measurement range and accuracy
Comen SpO
2
range: 0.05 - 20%; accuracy: not defined;
Masimo SpO
2
: 0.02% - 20%; accuracy: not defined;
PI Resolution
Comen SpO
2
:
0.05% - 9.99%, resolution: 0.01%,
10.0% - 20.0%, resolution: 0.1%.
Masimo SpO
2
:
0.02% - 9.99%, resolution: 0.01%.
10.0% - 20.0%, resolution: 0.1%.
Signal IQ (SIQ) indicator
Masimo SpO
2
and Comen SpO
2
modules have the SIQ feature.