Comen SpO
2
:
Measurement range: 20bpm-300bpm; resolution: 1bpm;
measurement error: ±2bpm.
Masimo SpO
2
:
Measurement range: 25bpm-240bpm; resolution: 1bpm;
measurement error: ±3bpm (non-motion) or ±5bpm (during motion).
Nellcor SpO
2
Measurement range: 20bpm-300bpm; resolution: 1bpm;
measurement error: ±3bpm in 20bpm~250bpm range, 251 bpm~300 bpm:
not defined
Comen NIBP:
Measurement range: 30bpm-300bpm; resolution: 1bpm;
measurement error: ±3bpm or ±3%, whichever is greater.
SunTec NIBP:
Measurement range: 30bpm-220bmp; resolution: 1bpm;
measurement error: ±3bpm or ±2%, whichever is greater.
IBP
Measurement range: 25bpm-350bpm; resolution: 1bpm;
measurement error: ±1bpm or ±1%, whichever is greater.
ADU mode: Upper limit: (lower limit+step size) bpm~300bpm;
lower limit: 15bpm~ (upper limit-step size) bpm.
PED/NEO mode: Upper limit: (lower limit+step size)bpm~300bpm;
lower limit: 15bpm~(upper limit-step size)bpm.