Product Specifications
II-9
10) PR specifications
Item
Specification
Measurement range and
accuracy
Comen SpO
2
:
Measurement range: 20bpm-300bpm; resolution: 1bpm;
measurement error: ±2bpm.
Masimo SpO
2
:
Measurement range: 25bpm-240bpm; resolution: 1bpm;
measurement error: ±3bpm (non-motion) or ±5bpm (during motion).
Nellcor SpO
2
Measurement range: 20bpm-300bpm; resolution: 1bpm;
measurement error: ±3bpm in 20bpm~250bpm range, 251 bpm300 bpm:
not defined
Comen NIBP:
Measurement range: 30bpm-300bpm; resolution: 1bpm;
measurement error: ±3bpm or ±3%, whichever is greater.
SunTec NIBP:
Measurement range: 30bpm-220bmp; resolution: 1bpm;
measurement error: ±3bpm or ±2%, whichever is greater.
IBP
Measurement range: 25bpm-350bpm; resolution: 1bpm;
measurement error: ±1bpm or ±1%, whichever is greater.
PR alarm settings
ADU mode: Upper limit: (lower limit+step size) bpm~300bpm;
lower limit: 15bpm~ (upper limit-step size) bpm.
PED/NEO mode: Upper limit: (lower limit+step size)bpm~300bpm;
lower limit: 15bpm~(upper limit-step size)bpm.
Step: 1540 bpm: 1bpm; 40300bpm: 5bpm
11) Temperature specifications
Name
Specification
The body temperature measurement meets the requirements of ISO 80601-2-56:2017+A1:2018
Measurement range and
accuracy
Measurement range
0°C -50°C (32°F-122°F )
Measurement error
±0.1°C 0.2°F ) (main unit)
Laboratory accuracy
±0.3°C (with temperature sensor error)
Alarm setup
Alarm limits
Upper limit: lower limit+0.1°C -50.0°C
Lower limit: 0°C ~upper limit-0.1
ΔT upper alarm limit
0.1°C -50.0°C
Display resolution
0.1°C (0.2°F )
Channel number
2 channels
Temperature sensor type
CY, YSI