System Alarm Messages
III-8
Check SpO
2
/SpO
2
b
Sensor Connection
Low
Not clearable
It needs to check the
connection of
SpO
2
/SpO
2
_2 sensor.
Please check the
connection of
SpO
2
/SpO
2
_2 sensor.
SpO
2
/SpO
2
b
Sensor Initializing
Prompt
/
SpO
2
/SpO
2
_2 Sensor
Initializing is
prompted.
/
Only SpO
2
/SpO
2
b
Reliable
Low
Not clearable
Only SpO
2
/SpO
2
_2 is
reliable
/
SpO
2
/SpO
2
b Demo
Mode
Prompt
/
Prompt SpO
2
/SpO
2
_2
Demo Mode is
displayed.
/
SpO
2
/SpO
2
b Finger
Sensor Off
High/Med,/Low
The alarm sound
and light are
clearable.
The SpO
2
Sensor is
disconnected with the
finger.
Check the condition of
SpO
2
sensor.
SpO
2
/SpO
2
b
Search Pulse
Timeout
Low
Not clearable
SpO
2
sensor is poorly
connected, or the
patient move his/her
arm.
Check the connection of
SpO
2
sensor.
SPO
2
/SpO
2
b
Sensor Fault
High
Not clearable
The sensor is fault.
Check and replace the
sensor; if the error still
exists, contact the
manufacturer for
maintenance.
SPO
2
/SpO
2
b
Module Error
High
Fully clearable
The module fails.
Contact the
manufacturer for
maintenance.
2.5) NIBP Technical Alarm
Alarm Message
Alarm Level
Alarm Clearance
Mode
Cause
Solution
NIBP Selftest Err.
Low
Fully clearable
An error occurs during
NIBP initialization.
Restart the system; if the
error still exists, contact
the manufacturer for
maintenance.
Cuff Loosen or
Disconnected
Low
Fully clearable
The NIBP cuff is poorly
connected.
Reconnect the NIBP cuff.
Pneumatic Leak
Low
Fully clearable
A leak exists in the
NIBP gas circuit.
Check the connection of
each part or replace the
cuff with a new one. If
the error still exists,
contact the
manufacturer for
maintenance.
Pressure
Overrange
Low
Fully clearable
An error occurs during
measuring the curve,
and the system cannot
perform the
measurement analysis
calculation.