System Alarm Messages
III-9
Cuff Inflation Tube
Leak
Low
Fully clearable
NIBP cuff is poorly
connected, or a leak
exists in the NIBP gas
circuit.
Air Pressure Error
Low
Fully clearable
An error occurs during
measuring the curve,
and the system cannot
perform the
measurement analysis
calculation.
Weak Signal
Low
Fully clearable
An error occurs during
measuring the curve,
and the system cannot
perform the
measurement analysis
calculation.
Check whether the
patient type is selected
correctly; Check the
connection of each part
replace the cuff with a
new one. If the error still
exists, contact the
manufacturer for
maintenance.
Arm Motion
Low
Fully clearable
The patient moves
his/her arm.
Check the connection of
each part and patient’s
condition, and measure
again. If the error still
exists, contact the
manufacturer for
maintenance.
Signal Saturated
Low
Fully clearable
An error occurs during
measuring the curve,
and the system cannot
perform the
measurement analysis
calculation.
NIBP System
Failure
High
Fully clearable
NIBP Measure
Timeout
Low
Fully clearable
NIBP Measure
Failed
Low
Fully clearable
Over Pressure
Low
Fully clearable
The gas circuit may be
folded or the module
fails.
Check whether the gas
circuit is blocked. Check
the patient's condition. If
the error still exists,
contact the
manufacturer for
maintenance.
NIBP Block
Low
Fully clearable
The discharge valve
may be blocked.
Check whether the
airway tube is bent
sharply or blocked; if the
error still exists, contact
the manufacturer for
maintenance.
NIBP Need
Calibrate
Prompt
/
NIBP Need Calibration
is prompted
/
Cuff Type Err.
Low
Fully clearable
The cuff type may not
match with the set
patient type.
Check whether the
patient type is selected
correctly; Check the