System Alarm Messages
III-10
connection of each part
or replace the cuff with a
new one. If the error still
exists, contact the
manufacturer for
maintenance.
NIBP Reset Error
Low
Fully clearable
An illegal reset is
performed during the
NIBP measurement.
Check whether a
blockage exists in the gas
circuit; then remeasure
it. If the error still exists,
contact the
manufacturer for
maintenance.
2.6) IBP Technical Alarm
Alarm Message
Alarm Level
Alarm Clearance
Mode
Cause
Solution
XX Sensor Off
Low
The alarm sound
and light are
clearable
The IBP sensor is
poorly connected.
Check or reinsert the IBP
sensor; if the error still
exists, contact the
manufacturer for
maintenance.
Zero Failed
Low
Not clearable
Zero IBP fails
Re-zero it.
Calib. Failed
Low
Not clearable
IBP Calib. Failed
Re-calibrate it.
Need Zero
Prompt
/
Need Zero is
prompted.
/
Sensor Fault
Med
Not clearable
Sensor Fault
Check the sensor and
replace it with a proper
one.
Zero Succeeded
Prompt
/
IBP ZERO SUCCESS
/
Calib. Succeeded
Prompt
/
BP CALI SUCCESS
/
XX Cable Shorted
Prompt
Remove the IBP cable
and sensor from
patient, or open the
cross valve to
atmosphere
Check the hydraulic
connection. Confirm the
cross valve open to
patient’s end. Replace
the cable and sensor if
zeroing fail again.
Note: XX represents a label name of IBP such as IBP: Art, IBP: PA, IBP: Ao, IBP: UAP, IBP: BAP, IBP: FAP, IBP: CVP, IBP:
LAP, IBP: RAP, IBP: ICP, IBP: UVP, IBP: LV, IBP: IAP, IBP: P1, IBP: P2, IBP: P3, IBP: P4, etc.
2.7) CO
2
Technical Alarm