System Alarm Messages
III-14
Alarm Message
Alarm Priority
Alarm Clearance
Mode
Possible cause
Solution
and the viewed bed is
disrupted.
Viewed by XX YY-
ZZ, disconnected
Low
Fully clearable
The connection
between monitor
being viewed and the
primary monitor is
disrupted.
Check the network
connection.
IP Address Conflict
Prompt
/
The IP address of
monitors are the
same.
Check the network
connection.
MAC Address
Conflict
Prompt
/
The MAC address of
monitors are the
same.
Check the network
connection.
2.10) Power Supply Technical Alarm
Alarm Message
Alarm Priority
Alarm Clearance
Mode
Possible cause
Solution
Battery Low
Medium
Not clearable
The battery capacity is
low.
Connect to mains supply
and charge the battery.
Battery Depletion
High
Not clearable
The battery is near
depletion and would
shut down.
Connect to mains supply
and charge the battery.
Power Board
Communication
Stop
High
Not clearable
Power board failure.
Restart the monitor. If
the error still exists,
contact the service
personnel for
maintenance.
Battery Fault
High
Not clearable
Battery error.
Change the battery.
2.11) Recorder Technical Alarm
Alarm Message
Alarm Priority
Alarm Clearance
Mode
Possible cause
Solution
Recorder is
printing
Prompt
/
The recorder is
printing.
/
Recorder thermal
array over
temperature
Low
Not clearable
The recorder is
printing for a long
time.
Stop printing and wait
until it cools down.
No Recorder
Paper
Low
The alarm sound
and light are
clearable
There is no recorder
paper or it is not
installed well.
Reload the recorder
paper.
No recorder
Prompt
/
No recorder
connected.
Please insert the
recorder module.