Default configuration information
IV-8
Work Mode
Measurement
O
2
Compensation
(Respironics/Comen)
16%
O
2
Compensation (Masimo)
Low
N
2
O Compensation (Masimo)
Off
Waveform Speed
6.25mm/s
Scale
50
Waveform Type
Line
Balance gas (Comen/ Respironics)
Room air
Altitude unit
m
Altitude(Comen, Respironics)
0m
Barometric Pressure
760mmHg
10. IBP default settings
Name
Default settings
IBP-S
Alarm On/Off
On
Alarm Limits (mmHg)
Art/AO/UAP/BAP/FAP/LV/P1-P4 arterial
pressure:
ADU: 90-160;
PED: 70-120;
NEO: 55-90
PA:
ADU: 10-35;
PED: 24-60;
NEO: 24-60
Alarm Level
Med
Alarm Auto Print
Off
IBP-D
Alarm On/Off
On
Alarm Limits (mmHg)
Art/AO/UAP/BAP/FAP/LV/P1-P4 arterial
pressure:
ADU: 50-90;
PED: 40-70
NEO: 20-60
PA:
ADU: 0-16;
PED: -4-4;
NEO: -4-4
Alarm Level
Med
Alarm Auto Print
Off
IBP-M
Alarm On/Off
On
Alarm Limits (mmHg)
Art/AO/UAP/BAP/FAP/LV/P1-P4 arterial
pressure:
ADU: 70-110;
PED: 50-90
NEO: 35-70