Default configuration information
IV-9
PA:
ADU: 0-20;
PED: 12-26;
NEO: 12-26
CVP/ICP/RAP/LAP/UVP/P1-P4 venous
pressure:
ADU: 0-10;
PED: 0-4
NEO: 0-4
Alarm Level
Med
Alarm Auto Print
Off
Art-S Extreme
Alarm On/Off
Off
Alarm Limits (mmHg)
ADU: 75-175
PED: 60-130
NEO: 50-95
Alarm Level
High
Alarm Auto Print
Off
Art-D Extreme
Alarm On/Off
Off
Alarm Limits (mmHg)
ADU: 35-105
PED: 30-80
NEO: 15-65
Alarm Level
High
Alarm Auto Print
Off
Art-M Extreme
Alarm On/Off
Off
Alarm Limits (mmHg)
ADU: 55-125
PED: 40-100
NEO: 30-75
Alarm Level
High
Alarm Auto Print
Off
CPP
Alarm On/Off
On
Alarm Limits (mmHg)
ADU: 50-130
PED: 40/-00
NEO: 30/-0
Alarm Level
Med
Alarm Auto Print
Off
Measure (P1, P2)
All
Measure (P3, P4)
Mean Only
Waveform Speed
25mm/s
PPV Measure
Off
Filter type
Normal
Scale
ART/AO/BAP/FAP/LV/P1/P2: 0~160
CVP/ICP/RAP/LAP/UVP/IAP: 0~20
UAP/P3/P4: 0~80
PA: 0~30
PAWP Timeout
15min