5 Human machine interface
Default parameter values
MB-20-G/MB-200G-W_MB-LABmaster_MB-200 MOD_TP700
Operating Manual - V5.0 - 04/2018 - STD
Page
5 - 15
Parameters for
MBRAUN basic systems
Factory
setting
Unit
Min
value
Max
value
Operator set
value
Min. Max.
Start circulation automat-
ically after regeneration
No
Regenerate automatically No
Regeneration interval 25 h 24 999
Box parameter
Box pressure: upper oper-
ating limit
+4.0 mbar -13.5 +14.5
Box pressure: lower oper-
ating limit
-4.0 mbar -14.5 +13.5
Hysteresis OAG 2.0 mbar 0.5 calc.
Hysteresis UAG 2.0 mbar 0.5 calc.
Box pressure: upper alarm
limit
+15.0 mbar -14.0 +15.0 Cannot be
changed
Box pressure: lower alarm
limit
-15.0 mbar -15.0 +14.0 Cannot be
changed
H
2
O alarm 20.0 ppm 5.0 499.0
O
2
alarm 20.0 ppm 5.0 999.0
Automatic purge of glovebox (optional)
Glovebox purge: autom.
Switch off after
60 min 0 999
Automatic purge if O
2
limit
value is exceeded
Yes
O
2
limit value 100.0 ppm 10.0 999.9
LMF parameters (optional)
Alarm threshold LMF ana-
lyser
100 ppm 100 999
Antechamber time controlled
Number of full antecham-
ber cycles
1 1 10
Time of evacuation 300 s 10 1000