5 Human machine interface
Default parameter values
Page
5 - 16
MB-20-G/MB-200G-W_MB-LABmaster_MB-200 MOD_TP700
Operating Manual - V5.0 - 04/2018 - STD
Parameters for
MBRAUN basic systems
Factory
setting
Unit
Min
value
Max
value
Operator set
value
Min. Max.
Refill time for intermediate
refilling
45 s 0 300
Refill time after the last
evacuation cycle
120 s 30 600
Antechamber pressure controlled (optional)
Threshold value for inter-
mediate refill
400 mbar 50 atmo-
sphere
Threshold value for va-
cuum leak test
5x10
-1
mbar 1x10
-2
10
Threshold value for final
vacuum
5x10
-1
mbar 1x10
-2
10
Number of pump refill
cycles
1 1 10
Maximum evacuation time 5 min 1 20
Maximum leakage rate
1
3 steps 1 10
Antechambers: lock inner
door after
2
60 min 0 999
1
Example: vacuum increases from 2 x10
-1
mbar to 4x10
-1
mbar (that is 2 steps). If this
is accurate, the process is stopped and the message “Antechamber leaks” is dis-
played.
2
Optional feature, refer to chapter "Antechamber operation"
Set value = 0 : function is deactivated