7 Gas purifier operation
Navigation overview
MB-20-G/MB-200G-W_MB-LABmaster_MB-200 MOD_TP700
Operating Manual - V5.0 - 04/2018 - STD
Page
7 - 3
Circulation LMF Filter 1: Switches the “Circulation over
LMF" function on and off.
Regenerat. LMF Filter 2: Switches the "Regeneration"
function on and off.
Vacuum Pump Solvent Trap: Switches the vacuum
pump, dedicated to the LMF, on and off.
Purification unit parameter screen
Start Screen ⇨ RKM icon ⇨ Parameter
► Touch the Para-
meter button.
ð The Purification Unit
Parameter screen is
displayed.
WARNING
Mismatch between parameter configuration and the regeneration gas (H
2
> 5%
⇔
H
2
< 5%)
Risk of fire in the disposal system or in the room
► Make sure that the set H
2
concentration in the Purification Unit Parameter screen and
the real H
2
concentration of regeneration gas matches.
► Make sure that the room, where the system is operated is well ventilated.