7 Gas purifier operation
Circulation
MB-20-G/MB-200G-W_MB-LABmaster_MB-200 MOD_TP700
Operating Manual - V5.0 - 04/2018 - STD
Page
7 - 9
Single purifier Functions screen
Start Screen Functions
Circulation mode is deac-
tivated. Regeneration and
box purge are enabled.
Touch the Circula-
tion Reactor 1 but-
ton.
ð Circulation is switched
on.
ð The regeneration
function is disabled.
ð The box purge func-
tion is disabled.
Start Screen Functions
Reactor1 is in circulation
mode and disabled for re-
generation. Function box
purge is disabled.
Touch the Circula-
tion Reactor 1 but-
ton.
ð Circulation is switched
off.
ð The box purge func-
tion is enabled.
ð The regeneration Re-
actor 1 function is en-
abled