7 Gas purifier operation
Regeneration operation
Page
7 - 22
MB-20-G/MB-200G-W_MB-LABmaster_MB-200 MOD_TP700
Operating Manual - V5.0 - 04/2018 - STD
Start Screen ⇨ RKM icon
Alternative:
► Touch the Regen-
erat. LMF Filter 2
button.
ð The button colour
changes from white to
green.
ð LMF2 changes to re-
generation status.
When regenerating sov-
ent filter 1, proceed ac-
cordingly.
Deactivating the regeneration
LMF regeneration is activated and deactivated in the same way as reactor regeneration.
Refer to Manual activation / deactivating of a reactor regeneration [}7 - 20].
Simultaneous regeneration of LMF with RKM
It is possible to regenerate an LMF and an RKM simultaneously:
¡ With additional vacuum pump (VPGL) for the solvent filter an easy simultaneous re-
generation is possible.
¡ With one vacuum pump used as shared resource:
¡ The regeneration started first has priority and runs according to programme.
¡ The regeneration started last, continues after confirmation of the regeneration gas
test in a regeneration step (step 2 of the regeneration programme). The next step
is started after a delay of approx. 10hours.