8B Freezer
Operation
MB-20-G/MB-200G-W_MB-LABmaster_MB-200 MOD_TP700
Operating Manual - V5.0 - 04/2018 - STD
Page
8B - 3
Parameters for
MBRAUN basic systems
Factory
setting
Unit
Min
value
Max
value
Operator set
value
Min. Max.
Freezer temp. target value -10 °C -40 +10
Freezer temp. alarm 0 °C -40 25
Glovebox cooling temp.
target value
20 °C 10 40
Glovebox cooling temp.
alarm
30 °C 10 50
In order to change the alarm value in the next menu:
► Touch the input field "Temperature alarm freezer".
ð The alphanumeric input screen opens (refer to the main documentation for further de-
tails).
► Confirm entries by touching the Enter button.
Freezer purge during glovebox purge
CAUTION
Opening a freezer, whose interior is filled with ambient atmosphere
Risk of damage to the process material!
opening a freezer when its interior is filled with ambient atmosphere, causes the box at-
mosphere to become polluted.
► Purge the interior of the freezer before using the glovebox.
During a glovebox purge, make sure that the interior of the freezer and the glovebox are
thoroughly purged and that no ambient atmosphere remains:
► When purging the glovebox, the freezer door must be fully opened.