12 Maintenance
Maintenance works
MB-20-G/MB-200G-W_MB-LABmaster_MB-200 MOD_TP700
Operating Manual - V5.0 - 04/2018 - STD
Page
12 - 15
► Insert the new glove into the old glove.
Take care that both gloves fit tightly into each
other so that there is no air between both
gloves.
► Remove both O-rings from the gloveport feed-
through.
► Remove the old glove from the gloveport and
immediately place the new glove over the port.
► Place the new glove over the port so that the
rim of the glove is in the port’s innermost
groove.
► Check that the glove is orientated correctly.
► Secure the new glove with one new O-ring at
the inner groove.
► Remove the old glove from inside.