12 Maintenance
Maintenance works
MB-20-G/MB-200G-W_MB-LABmaster_MB-200 MOD_TP700
Operating Manual - V5.0 - 04/2018 - STD
Page
12 - 15
Insert the new glove into the old glove.
Take care that both gloves fit tightly into each
other so that there is no air between both
gloves.
Remove both O-rings from the gloveport feed-
through.
Remove the old glove from the gloveport and
immediately place the new glove over the port.
Place the new glove over the port so that the
rim of the glove is in the port’s innermost
groove.
Check that the glove is orientated correctly.
Secure the new glove with one new O-ring at
the inner groove.
Remove the old glove from inside.