12 Maintenance
Maintenance works
Page
12 - 14
MB-20-G/MB-200G-W_MB-LABmaster_MB-200 MOD_TP700
Operating Manual - V5.0 - 04/2018 - STD
Gloveport feedthrough is sealed.
Roll up the new glove firmly so that as little air as pos-
sible remains in the new glove.
Place anatomical gloves in the correct position.
Place the new, rolled-up glove into the (sealed) glove
port feedthrough.
Put the new glove over the gloveport feedthrough and
insert the hardened rim of the cuff into the inner
notch.
Secure the glove by putting two new O-rings onto the
outer two notches of the gloveport feedthrough.
Remove the inner gloveport cover.
Glove change without gloveport cover
Create a slight overpressure in the box (approx. 1-2 mbar).