4 Transport, installation and commissioning
Installation
MB-20-G/MB-200G-W_MB-LABmaster_MB-200 MOD_TP700
Operating Manual - V5.0 - 04/2018 - STD
Page
4 - 13
¡ If an (effective) endless gas source is used for an automatic or a manual box
purge, the identical conditions (refer to Preparing the site [}4 - 13]) have to be
applied as they are applied in cases, the conditions of the site location are not
met.
Risk of O
2
concentration falling below minimum (18%)
If the room size requirements of the site location are not met or an endless gas source is
used, there is a risk of the O
2
concentration to fall below the lowest allowed level.
► In such locations attach a note at the room door, e.g. “In case of a glove tear during
glovebox operation, the room is allowed to be entered only wearing an oxygen
mask”.
► Alternatively: an additional security kit is available. It stops the gas supply through a
redundant interlock system. It can be ordered at MBRAUN.
► Additionally, MBRAUN recommends securing the site location with a gas monitor
that triggers an alarm if the oxygen concentration falls below the required minimum
level.
► The concentration alarm allows entering the room only wearing an oxygen mask. If
the alarm is on, only experienced and skilled personnel are allowed to enter the
room.
► If in doubt, contact MBRAUN Service for assistance.
Installation of system components in different locations
If the gas purifier and the glovebox are installed in different locations, the instructions out-
lined above apply to both components.
If you have any questions, contact MBRAUN Service.