R&S FSL CALCulate Subsystem (Analog Demodulation, K7)
1300.2519.12 6.293 E-11
C
ALCulate<1|2>:MARKer:FUNCtion:ADEMod:SINad:RESult<1...6>?
This command queries the result of the SINAD measurement. The numeric suffix
(:RESult<1...6>) indicates whether trace 1, 2, 3, 4, 5 or 6 is selected. The numeric suffixes
<1|2> are irrelevant.
This command is a query only and thus has no *RST value.
Example
ADEM ON
Switches on analog demodulator
CALC:FEED 'XTIM:FM:AFSP'
DISP:TRAC ON
Switches on AF spectrum of FM and trace
CALC:MARK:FUNC:ADEM:SIN:RES?
Queries SINAD value
Characteristics
*RST value: –
SCPI: device–specific
Mode
ADEMOD
CALCulate<1|2>:MARKer:FUNCtion:ADEMod:THD:RESult<1...6>?
This command queries the result of the THD measurement. The numeric suffix
(:RESult<1...6>) indicates whether trace 1, 2, 3, 4, 5 or 6 is selected. The numeric suffixes
<1|2> are irrelevant.
This command is a query only and thus has no *RST value.
Example
ADEM ON
Switches on analog demodulator
CALC:FEED 'XTIM:FM:AFSP'
Switches on AF spectrum of FM and trace
DISP:TRAC ON
Switches on the trace
CALC:MARK:FUNC:ADEM:THD:RES?
Queries THD result
Characteristics
*RST value: –
SCPI: device–specific
Mode
ADEMOD