SENSe Subsystem (Analog Demodulation, K7) R&S FSL
1300.2519.12 6.304 E-11
[
SENSe:]ADEMod:AF:STOP
This command sets the stop frequency for AF spectrum result display.
T
he stop frequency is limited to half the measurement bandwidth of analog demodulation
([SENSe:]ADEMod:BANDwidth|BWIDth:DEModulation).
Parameter
<numeric_value>
Example
ADEM ON
Switches on the analog demodulator
CALC:FEED 'XTIM:FM:AFSP'
Switches on AF spectrum result display of FM
ADEM:BAND 5 MHz
Sets the measurement bandwidth to 5 MHz
ADEM:AF:STAR 0kHz
Sets the AF start frequency to 0 kHz
ADEM:AF:STOP 500kHz
Sets the AF stop frequency to 500 kHz
Characteristics
*RST value: 2.5 MHz
SCPI: device–specific
Mode
ADEMOD
[SENSe:]ADEMod:AM[:ABSolute][:TDOMain][:TYPE]
This command selects the result types of the RF signal to be measured simultaneously in zero
span.
Parameter
<result type 1|2|3|4|5|6>: WRITe, AVERage, MAXHold, MINHold, VIEW, OFF; for details see
"Trace mode result types" on page 6.300.
Example
ADEM:AM AVER,MAXH,MINH
Determines average, max hold and min hold values at a time.
ADEM:AM WRIT,OFF,OFF
Determines only the current measurement values.
ADEM:AM OFF,OFF,OFF
Switches AM demodulation off.
Characteristics
*RST value: WRITe,OFF,OFF
SCPI: device–specific
Mode
ADEMOD