EasyManua.ls Logo

Rohde & Schwarz R&S FSL3 - Page 1063

Rohde & Schwarz R&S FSL3
1734 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SENSe Subsystem (Analog Demodulation, K7) R&S FSL
1300.2519.12 6.304 E-11
[
SENSe:]ADEMod:AF:STOP
This command sets the stop frequency for AF spectrum result display.
T
he stop frequency is limited to half the measurement bandwidth of analog demodulation
([SENSe:]ADEMod:BANDwidth|BWIDth:DEModulation).
Parameter
<numeric_value>
Example
ADEM ON
Switches on the analog demodulator
CALC:FEED 'XTIM:FM:AFSP'
Switches on AF spectrum result display of FM
ADEM:BAND 5 MHz
Sets the measurement bandwidth to 5 MHz
ADEM:AF:STAR 0kHz
Sets the AF start frequency to 0 kHz
ADEM:AF:STOP 500kHz
Sets the AF stop frequency to 500 kHz
Characteristics
*RST value: 2.5 MHz
SCPI: device–specific
Mode
ADEMOD
[SENSe:]ADEMod:AM[:ABSolute][:TDOMain][:TYPE]
This command selects the result types of the RF signal to be measured simultaneously in zero
span.
Parameter
<result type 1|2|3|4|5|6>: WRITe, AVERage, MAXHold, MINHold, VIEW, OFF; for details see
"Trace mode result types" on page 6.300.
Example
ADEM:AM AVER,MAXH,MINH
Determines average, max hold and min hold values at a time.
ADEM:AM WRIT,OFF,OFF
Determines only the current measurement values.
ADEM:AM OFF,OFF,OFF
Switches AM demodulation off.
Characteristics
*RST value: WRITe,OFF,OFF
SCPI: device–specific
Mode
ADEMOD