SENSe Subsystem (Analog Demodulation, K7) R&S FSL
1300.2519.12 6.314 E-11
ADEM:PM:RES? AVER
Reads the PM average result data.
ADEM:PM:RES? MAXH
Reads the PM maxhold result data.
ADEM:PM:RES? MINH
Reads the PM minhold result data.
ADEM:AM:RES? WRIT
Reads the current AM result data.
Characteristics
RST value:
SCPI: device–specific
Mode
ADEMOD
[SENSe:]ADEMod:PM:AFSPectrum[:TYPE]
This command selects the AF spectrum result types of the PM–demodulated signal to be
measured simultaneously.
Parameter
<result type 1|2|3|4|5|6>: WRITe, AVERage, MAXHold, MINHold, VIEW; for details see "Trace
mode result types" on page 6.300.
Note: The result type "AF spectrum of the PM demodulated signal" cannot be activated at the
same time as "AF spectrum of AM or FM demodulated signal".
Example
ADEM:PM:AFSP AVER,MAXH,MINH
Determines average, maximum and minimum value simultaneously
ADEM:PM:AFSP WRIT,OFF,OFF
Determines only current measurement results
ADEM:PM:AFSP OFF,OFF,OFF
Switches calculation of AF spectrum off
Characteristics
RST value: OFF,OFF,OFF
SCPI: device–specific
Mode
ADEMOD