R&S FSL SENSe Subsystem (Analog Demodulation, K7)
1300.2519.12 6.313 E-11
[
SENSe:]ADEMod:PM[:TDOMain][:TYPE]
This command selects the result types of the PM–demodulated signal to be created
simultaneously.
Parameter
<result type 1|2|3|4|5|6>: WRITe, AVERage, MAXHold, MINHold, VIEW; for details see "Trace
mode result types" on page 6.300.
Example
ADEM:PM AVER,MAXH,MINH
Determines average, maximum and minimum value simultaneously
ADEM:PM WRIT,OFF,OFF
Determines only current measurement results
ADEM:PM OFF,OFF,OFF
Switches the PM demodulator off.
Characteristics
RST value: OFF,OFF,OFF
SCPI: device–specific
Mode
ADEMOD
[SENSe:]ADEMod:PM[:TDOMain]:RESult?
This command reads the result data of the PM demodulation for the specified result type. The
data format of the output data is determined with the FORMat command.
Parameter
<result type>: WRITe, AVERage, MAXHold, MINHold; for details see "Trace mode result types"
on page 6.300.
Note: The specified result type must be one of those configured with the
[SENSe:]ADEMod:PM[:TDOMain][:TYPE] command. Otherwise a query error will
be generated.
Return values
ASCII Format (FORMat ASCII) or Binary Format (FORMat REAL,32); for details see "Formats
for returned values: ASCII format and binary format" on page 6.300. The ouput units are
described in "CALCulate<1|2>:MARKer<1...4>:PEXCursion" on page 6.63.
Example
ADEM:SET 8MHz,32000,EXT,POS,–500,30
Sets the demodulator parameters.
ADEM:PM AVER,MAXH,MINH
Selects the PM results to be measured.
ADEM:AM WRIT,OFF,OFF
Selects the AM results to be measured.
ADEM ON
Switches on the demodulator.
INIT;*WAI
Starts the measurement and waits for termination.
FORM ASC
Selects the output format.