R&S FSL SENSe Subsystem (Analog Demodulation, K7)
1300.2519.12 6.315 E-11
[
SENSe:]ADEMod:PM:AFSPectrum:RESult?
This command reads out the AF spectrum result data of the PM–demodulated signal for the specified
result type. The data format of the output data is determined with the FORMat command.
P
arameter
<result type>: WRITe, AVERage, MAXHold, MINHold; for details see "Trace mode result types"
on page 6.300.
Note: The specified result type must be one of those configured with the
[SENSe:]ADEMod:PM:AFSPectrum[:TYPE] command. Otherwise a query error will
be generated.
Return values
ASCII Format (FORMat ASCII) or Binary Format (FORMat REAL,32); for details see "Formats
for returned values: ASCII format and binary format" on page 6.300. The ouput units are
described in section "CALCulate<1|2>:MARKer<1...4>:PEXCursion" on page 6.63.
Example
ADEM:SET 8MHz,32000,EXT,POS,–500,30
Sets demodulator
ADEM:PM AVER,MAXH,MINH
Selects the PM results to be measured
ADEM:AM:REL WRIT,OFF,OFF
Selects the AM results to be measured
ADEM:PM:AFSP WRIT,OFF,OFF
Selects the AF spectrum results of the demodulated PM signal to be measured
ADEM ON
Switches the demodulator on
INIT;*WAI
Starts the measurement and waits for termination
FORM ASC
Selects output format
ADEM:PM:RES? AVER
Reads PM average result data
ADEM:PM:RES? MAXH
Reads PM maxhold result data
ADEM:PM:RES? MINH
Reads PM minhold result data
ADEM:AM:RES? WRIT
Reads current AM result data
ADEM:PM:AFSP:RES? WRIT
Reads current AF spectrum result data of demodulated PM signal
Characteristics
RST value: –
SCPI: device–specific
Mode
ADEMOD