R&S FSL SENSe Subsystem (Analog Demodulation, K7)
1300.2519.12 6.323 E-11
SENSe:BANDwidth Subsystem (Analog Demodulation, K7)
This subsystem controls the setting of the instruments filter bandwidths. Both groups of
commands (BANDwidth and BWIDth) perform the same functions.
Commands of the SENSe:BANDwidth Subsystem
– [SENSe<1|2>:]BANDwidth|BWIDth:DEMod
[SENSe<1|2>:]BANDwidth|BWIDth:DEMod
This command sets the bandwidth for analog demodulation. Depending on the selected
demodulation bandwidth, the instrument selects the required sampling rate.
The available values of the demodulation bandwidths are determined by the sampling rates. For
details on the correlation between demodulation bandwidth and sampling rate refer to chapter
"Instrument Functions", sections "Analog Demodulation (Option K7)" – "Sample rate,
measurement time and trigger offset".
Parameter
<numeric_value>
Example:
BAND:DEM 1MHz
Sets test bandwidth to 1 MHz
Characteristics:
*RST value: 5 MHz
SCPI: device–specific
Mode
A–F, ADEMOD