EasyManua.ls Logo

Rohde & Schwarz R&S FSL3 - Page 1102

Rohde & Schwarz R&S FSL3
1734 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S FSL CALCulate:BTOoth Subsystem (BLUETOOTH, K8)
1300.2519.12 6.343 E-11
Characteristics
RST value:
SCPI: device–specific
Mode
BT
CALCulate<1|2>:BTOoth:IBSemissions:EXCeptions?
This command reads the number of results that exceeded the specified limits in the In–band
Spurious Emissions measurement.
The numeric suffixes <1|2> are irrelevant for this command.
Note: This command is only available with active In–band Spurious Emissions measurement
(see CONFigure:BTOoth:MEASurement command). With all other measurements
this command will lead to a query error.
Example
For the first steps refer to "Example: Adapting the settings to the characteristics of the DUT" on
page 6.334.
CONF:BTO:MEAS IBS
Activates the In–band Spurious Emissions measurement.
CONF:BTO:CHAN 7
Adjust the TX channel number.
INIT;*WAI
Starts the measurement with synchronization.
CALC:BTO:IBS:EXC?
Reads the number of results that exceeded the specified limits.
Characteristics
RST value:
SCPI: device–specific
Mode
BT
CALCulate<1|2>:BTOoth:IBSemissions:HADJacent?
This command reads the maximum power of the upper or lower frequency band (500 kHz) of
the TX channel in the In–band Spurious Emissions measurement.
The numeric suffixes <1|2> are irrelevant for this command.
Note: This command is only available with active In–band Spurious Emissions measurement
(see CONFigure:BTOoth:MEASurement command). With all other measurements
this command will lead to a query error.
Parameter
UPPer | LOWer

Table of Contents