CALCulate:BTOoth Subsystem (BLUETOOTH, K8) R&S FSL
1300.2519.12 6.344 E-11
E
xample
For the first steps refer to "Example: Adapting the settings to the characteristics of the DUT" on
page 6.334.
CONF:BTO:MEAS IBS
Activates the In–band Spurious Emissions measurement.
CONF:BTO:CHAN 7
Adjust the TX channel number.
INIT;*WAI
Starts the measurement with synchronization.
CALC:BTO:IBS:HADJ? LOW
Reads the maximum power of the lower frequency band.
Characteristics
RST value:
SCPI: device–specific
Mode
BT
CALCulate<1|2>:BTOoth:IBSemissions:TXReference?
This command reads the TX channel reference power.
The numeric suffixes <1|2> are irrelevant for this command.
Note: This command is only available with active In–band Spurious Emissions measurement
(see CONFigure:BTOoth:MEASurement command). With all other measurements
this command will lead to a query error.
Example
For the first steps refer to "Example: Adapting the settings to the characteristics of the DUT" on
page 6.334.
CONF:BTO:MEAS IBS
Activates the In–band Spurious Emissions measurement.
CONF:BTO:CHAN 7
Adjust the TX channel number.
INIT;*WAI
Starts the measurement with synchronization.
CALC:BTO:IBS:TXR?
Reads the TX channel reference power.
Characteristics