R&S FSL CALCulate Subsystem (CATV, K20)
1300.2519.12 6.455 E-11
C
ALCulate:DTV:LIMit:RESult:MERPeak:UPPer?
This command performs an upper limit check for the modulation error rate peak in the Overview
measurement.
This command is a query and therefore has no *RST value.
Example
CALC:DTV:LIM:MERP:UPP 20
Sets the upper limit for the modulation error rate peak to 20%.
INIT;*WAI
Starts a new measurement and waits for the sweep end.
CALC:DTV:LIM:RES:MERP:UPP?
Performs an upper limit check for the modulation error rate peak.
Characteristics
RST value: –
SCPI: device–specific
Mode
CATV
CALCulate:DTV:LIMit:RESult:MERRms:LOWer?
This command performs a lower limit check for the root mean square value of the modulation
error rate in the Overview measurement.
This command is a query and therefore has no *RST value.
Example
CALC:DTV:LIM:MERR:LOW 30
Sets the lower limit for the root mean square value of the modulation error rate to 30 dB.
INIT;*WAI
Starts a new measurement and waits for the sweep end.
CALC:DTV:LIM:RES:MERR:LOW?
Performs a lower limit check for the root mean square value of the modulation error rate.
Characteristics
RST value: –
SCPI: device–specific
Mode
CATV
CALCulate:DTV:LIMit:RESult:MERRms:UPPer?
This command performs an upper limit check for the root mean square value of the modulation
error rate in the Overview measurement.
This command is a query and therefore has no *RST value.
Example
CALC:DTV:LIM:MERR:UPP 30
Sets the upper limit for the root mean square value of the modulation error rate to 30%.
INIT;*WAI
Starts a new measurement and waits for the sweep end.