SENSe Subsystem (CATV, K20) R&S FSL
1300.2519.12 6.492 E-11
[
SENSe<1|2>:]ATV:CSO:TABle<1...10>:MFRequencies
This command defines the measurement frequencies (center frequencies and span values), and
activates or deactivates them for the CSO measurement.
The numeric suffixes <1|2> are irrelevant.
The numeric suffix <1...10> defines the table row.
Parameter
0 | 1,<numeric_value>,<numeric_value>
0 deactivated
1 activated
first numeric value center frequency in Hz
second numeric value span in Hz
Example
ATV:CSO:TABL1:MFR 1,–1000,1000
Activates the entries of the first table row for the measurement and sets the center frequency to
–1 kHz and the span to 1 kHz.
ATV:CSO:TABL2:MFR 1,–2000,2000
Activates the second row entries for the measurement and sets the center frequency to –2 kHz
and the span to 2 kHz.
Characteristics
RST value:
Row State Center frequency Span
1 ON –1.25 MHz 200 kHz
2 ON –750 kHz 200 kHz
3 ON –250 kHz 200 kHz
4 ON 250 kHz 200 kHz
5 ON 750 kHz 200 kHz
6 ON 1.25 MHz 200 kHz
7 ON 2.25 MHz 200 kHz
8 OFF 3.25 MHz 200 kHz
9 OFF 4.25 MHz 200 kHz
10 OFF 5.25 MHz 200 kHz
SCPI: device–specific
Mode
CATV