R&S FSL SENSe Subsystem (CATV, K20)
1300.2519.12 6.493 E-11
SENSe:ATV:CTB Subsystem (CATV, K20)
The SENSe:ATV:CTB subsystem contains commands for the Carrier–to–Composite Triple Beat Ratio
(CTB) measurement setup.
Commands of the SENSe:ATV:CTB Subsystem
– [SENSe<1|2>:]ATV:CTB:CFRelative:NEXT
– [SENSe<1|2>:]ATV:CTB:POWer:NCORrection
– [SENSe<1|2>:]ATV:CTB:POWer:REFerence:CHANnel:MANual
– [SENSe<1|2>:]ATV:CTB:POWer:REFerence:MANual
– [SENSe<1|2>:]ATV:CTB:POWer:REFerence:MODE
– [SENSe<1|2>:]ATV:CTB:TABle<1...10>:MFRequencies
[SENSe<1|2>:]ATV:CTB:CFRelative:NEXT
This command switches from one frequency range to the next according to the measurement
frequencies defined by the [SENSe<1|2>:]ATV:CTB:TABle<1...10>:MFRequencies command.
The numeric suffixes <1|2> are irrelevant.
This command is an event and therefore has no *RST value and no query.
Example
ATV:CTB:TABL1:MFR 1,100,10000
Activates the entries of the first table row for the measurement and sets the center frequency to
100 Hz and the span to 10 kHz.
ATV:CTB:TABL2:MFR 1,200,20000
Activates the second row entries for the measurement and sets the center frequency to 200 Hz
and the span to 20 kHz.
ATV:CTB:CFR:NEXT
Switches from the frequency range defined in table row 1 to 2.
Characteristics
RST value: –
SCPI: device–specific
Mode
CATV
[SENSe<1|2>:]ATV:CTB:POWer:NCORrection
This command activates or deactivates the noise floor correction.
The numeric suffixes <1|2> are irrelevant.
Parameter
ON | OFF
Example
ATV:CTB:POW:NCOR ON
Activates the noise floor correction.