SENSe Subsystem (CATV, K20) R&S FSL
1300.2519.12 6.494 E-11
C
haracteristics
RST value: OFF
SCPI: device–specific
M
ode
CATV
[SENSe<1|2>:]ATV:CTB:POWer:REFerence:CHANnel:MANual
This command defines the reference channel.
The numeric suffixes <1|2> are irrelevant.
Parameter
<numeric_value>, corresponding to active channel
Example
ATV:CTB:POW:REF:CHAN:MAN 5
Sets channel 5 in the active channel table as reference channel.
Characteristics
RST value: active measurement channel
SCPI: device–specific
Mode
CATV
[SENSe<1|2>:]ATV:CTB:POWer:REFerence:MANual
This command defines the manual reference power. The unit is set via the
CALCulate<1|2>:UNIT:POWer command.
The numeric suffixes <1|2> are irrelevant.
Parameter
–200 to 200 dB
Example
ATV:CTB:POW:REF:MAN 10
Sets the reference power to 10.
Characteristics
RST value: 0
SCPI: device–specific
Mode
CATV