R&S FSL SENSe Subsystem (CATV, K20)
1300.2519.12 6.497 E-11
SENSe:CORRection Subsystem (CATV, K20)
The SENSe:CORRection subsystem controls correction factors.
Commands of the SENSe:CORRection Subsystem
[SENSe<1|2>:]CORRection:RVELocity
[SENSe<1|2>:]CORRection:RVELocity
This command defines the velocity of propagation of the signal in the transmission channel (e.g.
cable), relative to speed of light.
The numeric suffixes <1|2> are irrelevant.
Parameter
0.0001 to 1.0
Example
CONF:DTV:MEAS EPAT
Configures the R&S FSL for the Echo Pattern measurement.
CORR:RVEL 0.2
Sets the velocity of propagation to 0.2.
Characteristics
RST value: 1
SCPI: device–specific