Loading...
R&S FSL CALCulate Subsystem (Noise Figure, K30)
1300.2519.12 6.525 E-11
C
ALCulate<1|2>:LIMit<1...6>:TRACe
This command assigns a limit line to a particular measurement type.
T
he numeric suffixes <1...6> indicate the limit line.
Parameter
NFIGure noise figure
TEFFective noise temperature
GAIN gain
Example
CALC:LIM2:TRAC NFIG
Assigns limit line 2 to the noise figure measurement.
CALC:LIM3:TRAC GAIN
Assigns limit line 3 to the gain measurement.
Characteristics
*RST value: –
SCPI: device–specific