EasyManua.ls Logo

Rohde & Schwarz R&S FSL3 - Page 1283

Rohde & Schwarz R&S FSL3
1734 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CALCulate Subsystem (Noise Figure, K30) R&S FSL
1300.2519.12 6.524 E-11
CALCulate<1|2>:LIMit<1...6>:COPY
Parameter
1 to 6 number of the new limit line
<name> name of the new limit line given as a string (max. 8 alphanumeric characters)
For further details refer to "CALCulate<1|2>:LIMit<1...8>:COPY" on page 6.30.
CALCulate<1|2>:LIMit<1...6>:DELete
Different to the base unit, only 6 limit lines can be active at the same time (indicated by LIMIT1
to LIMIT6).
For further details refer to "CALCulate<1|2>:LIMit<1...8>:DELete" on page 6.30.
CALCulate<1|2>:LIMit<1...6>:FAIL?
Different to the base unit, only 6 limit lines can be active at the same time (indicated by LIMIT1
to LIMIT6).
For further details refer to "CALCulate<1|2>:LIMit<1...8>:FAIL?" on page 6.31.
CALCulate<1|2>:LIMit<1...6>:NAME
This command assigns a name to a limit line. If it does not exist already, a limit line with this
name is created.
The numeric suffixes <1...6> indicate the limit line.
Parameter
<name of limit line>, max. 8 alphanumeric characters
Example
CALC:LIM1:NAME FM1
Assigns the name FM1 to limit line 1.
Characteristics
*RST value: REM1 to REM6 for lines 1 to 6
SCPI: device–specific
Mode
NF
CALCulate<1|2>:LIMit<1...6>:STATe
Different to the base unit, only 6 limit lines can be active at the same time (indicated by LIMIT1
to LIMIT6).
For further details refer to "CALCulate<1|2>:LIMit<1...8>:STATe" on page 6.31.