CALCulate Subsystem R&S FSL
1300.2519.12 6.30 E-11
C
haracteristics
*RST value: blank comment
SCPI: device–specific
M
ode
A, ADEMOD, CDMA, EVDO, NF
CALCulate<1|2>:LIMit<1...8>:COPY
This command copies one limit line onto another one
The numeric suffixes <1|2> are irrelevant for this command.
This command is an event and therefore has no *RST value and no query.
This command is available from firmware version 1.10.
Parameter
1 to 8 number of the new limit line
<name> name of the new limit line given as a string
Example
CALC:LIM1:COPY 2
Copies limit line 1 to line 2.
CALC:LIM1:COPY 'FM2'
Copies limit line 1 to a new line named FM2.
Characteristics
*RST value: –
SCPI: device–specific
Mode
A, ADEMOD, CDMA, EVDO, NF
CALCulate<1|2>:LIMit<1...8>:DELete
This command deletes the selected limit line.
The numeric suffixes <1|2> are irrelevant for this command.
This command is an event and therefore has no *RST value and no query.
This command is available from firmware version 1.10.
Example
CALC:LIM1:DEL
Deletes limit line 1.
Characteristics
*RST value: –
SCPI: device–specific
Mode
A, ADEMOD, CDMA, EVDO, NF