R&S FSL SENSe Subsystem (Noise Figure, K30)
1300.2519.12 6.549 E-11
[SENSe:]CORRection:ENR:MODE
This command specifies whether a constant ENR value applies for all measured frequencies or
an ENR table (ENR values specified at specific input frequencies). The constant ENR value is
specified by the [SENSe:]CORRection:ENR:SPOT command.
Parameter
TABLe ENR table is used.
SPOT Constant ENR value is used.
Example
CORR:ENR:MODE SPOT
Uses the configured constant ENR value for all measured frequencies.
Characteristics
*RST value: SPOT
SCPI: device–specific
Mode
NF
[SENSe:]CORRection:ENR:SPOT
This command sets the constant ENR value of the noise source that is used throughout the
entire frequency range.
Parameter
–999.99 to 999.99 dB
Example
CORR:ENR:MODE SPOT
Uses the configured constant ENR value for all measured frequencies.
CORR:ENR:SPOT 30
Sets the constant ENR value to 30 dB for all input frequencies to be measured.
Characteristics
*RST value: 15 dB
SCPI: device–specific
Mode
NF
[SENSe:]CORRection:ENR[:MEASurement]:TABLe:DATA
This command specifies a new ENR table to determine the correct ENR (excess noise ratio)
figure to be used for the input frequencies. This new list completely overwrites all current ENR
f re qu en c y li st entri e s r eg ar dle ss of h o w m any e ntries a r e pr esent a nd h ow m a n y entr ie s are bein g
s u ppl i e d for the ne w l is t .
Parameter
<numeric_value>,<numeric_value>, ... (maximum of 100 argument pairs)
numeric value = a frequency ENR pair of arguments
frequency: 0 Hz to 999.99 GHz.
ENR figure: –999.99 to 999.99 (value in dB, accurate to two decimal places)