SENSe Subsystem (Noise Figure, K30) R&S FSL
1300.2519.12 6.550 E-11
Example
CORR:ENR:MODE TABL
Uses the ENR table.
CORR:ENR:MEAS:TABL:DATA 1MHZ,10,2MHZ,12
Specifies a new ENR table and overwrites the current ENR table with the two entry pairs
specified.
C
haracteristics
*RST value: –
SCPI: device–specific
Mode
NF
[SENSe:]CORRection:IREJection
This command specifies an image rejection value for the selected DUT
([SENSe:]CONFigure:MODE:DUT command) which will be effective throughout the entire
frequency range.
Parameter
0 dB to 999.99 dB
Example
CONF:MODE:DUT DOWN
The DUT converts the input frequency to a lower output frequency.
CORR:IREJ 100
Sets an image rejection value of 100 dB.
Characteristics
*RST value: 999.99 dB
SCPI: device–specific
Mode
NF
[SENSe:]CORRection:LOSS:INPut:MODE
This command specifies whether a constant input loss value applies for all measured
frequencies or an input loss list (loss input values specified at specific input frequencies). The
constant loss input value is specified by the [SENSe:]CORRection:LOSS:INPut:SPOT
command.
Parameter
SPOT The constant loss input value for all measurement frequencies is used.
TABLe The loss input table is used.
Example
CORR:LOSS:INP:MODE SPOT
Uses the configured constant loss input value for all input frequencies to be measured.