SENSe Subsystem (Noise Figure, K30) R&S FSL
1300.2519.12 6.554 E-11
Measurement Settings
Commands of the measurement settings
– [SENSe:]BANDwidth|BWIDth[:RESolution]
– [SENSe:]CONFigure:MODE:DUT
– [SENSe:]CONFigure:MODE:SYSTem:LOSCillator:FREQuency
– [SENSe:]FREQuency[:CW|:FIXed]
– [SENSe:]FREQuency:LIST:DATA
– [SENSe:]FREQuency:STARt
– [SENSe:]FREQuency:STEP
– [SENSe:]FREQuency:STOP
– [SENSe:]SWEep:COUNt
– [SENSe:]SWEep:TIME
[SENSe:]BANDwidth|BWIDth[:RESolution]
For details refer to "[SENSe<1|2>:]BANDwidth|BWIDth[:RESolution]" on page 6.175.
[SENSe:]CONFigure:MODE:DUT
This command defines the type of DUT. This setting determines the method that is used to
create the frequency measurement list. Fixed local oscillator frequencies are taken from the
settings supplied by [SENSe:]CONFigure:MODE:SYSTem:LOSCillator:FREQuency.
Parameter
AMPLifier The DUT is an amplifier and not a frequency converting device.
DOWNconv The DUT converts the input frequency to a lower output frequency:
Fixed LO, IF=RF+LO
UPConv The DUT converts the input frequency to a higher output frequency:
Fixed LO, IF=abs(RF–LO).
Example
SENS:CONF:MODE:DUT DOWN
The DUT converts the input frequency to a lower output frequency.
Characteristics
*RST value: AMPLifier
SCPI: device–specific
Mode
NF